Próteses ao ecocardiograma

Figure2

.

https://www.hvt-journal.com/articles/art46

.

Table 3. Echocardiographic indices of the severe obstruction of replaced aortic valve (7, 8, 13, 15, 47)
Thickening or limited mobility of cusps or occluding discs
Calculated by Doppler method effective orifice area (EOA)  >0.35 cm2 less than expected  (in comparison with normal values for implanted type of prosthesis)
Narrowed systolic colour flow through the valve
Change in measurements more than 25% on serial studies
In a case of the previous data absence:

  • Effective orifice area (EOA) <0.8 cm2
  • Peak velocity (V max) ≥4.0 ms
  • Mean pressure gradient ≥35 mm Hg
  • Acceleration time of systolic flow >100 ms

.

Artigo muito bom para quem ficou curioso com a aula do DIC.

.

  • Acceleration time of systolic flow >100 ms

Surpresa vermos o uso desta importante informação ao ecocardiograma que têm sido menosprezada nos últimos tempos.

.

A aceleração prolongada é um excelente marcado da resistência ao esvaziamento ventricular e marcador de estenose funcionalmente significativa.

.

 

 