Because the formula for the cross-sectional area involves the square of the radius, any error in measuring the left ventricular outflow tract may create a substantial error in flow calculation. A 2-mm error in measuring a 2.0-cm diameter outflow tract will result in an approximate 20% error in the flow volume calculation.
.
.