Table 3. Echocardiographic indices of the severe obstruction of replaced aortic valve (7, 8, 13, 15, 47)
Thickening or limited mobility of cusps or occluding discs
Calculated by Doppler method effective orifice area (EOA)  >0.35 cm2 less than expected  (in comparison with normal values for implanted type of prosthesis)
Narrowed systolic colour flow through the valve
Change in measurements more than 25% on serial studies
In a case of the previous data absence:

  • Effective orifice area (EOA) <0.8 cm2
  • Peak velocity (V max) ≥4.0 ms
  • Mean pressure gradient ≥35 mm Hg
  • Acceleration time of systolic flow >100 ms

.